Recessed Through Wall Hospital Sterilizer Equipment With Pre-vacuum And Post Vacuum Phase
Specifications
Model No | Doors | Heating Type | Power Supply | Volume | Internal Chamber Dimension | Overall Dimension |
(Liters) | (w x h x d) | (w x h x d) | ||||
MD-0.8Y | Sliding door | External steam supply | AC380V, 50Hz, 3P/N/E | 800 | 680*1180*1000 | 2200*2000*1410 |
Outside facilities needed | ||||||
Electric power:AC380,50HZ,3 phase/N/E | ||||||
Steam feeding: Saturated steam from external steam pipes 0.4-0.7Mpa | ||||||
Water: Softened water, hardness <50ppm, pH Neutral, 0.15-0.3Mpa | ||||||
Compressed air: Oil free air, 0.5-0.8Mpa |
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: ●Packaged instruments ●Empty glass containers ●Empty metal containers Porous materials to be sterilized: ●Rubber articles |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced ●Textile |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |
Description
This series hospital sterilizer operate with saturated steam as the sterilizing agent with a
temperature range of 105 °C (221°F) to 134 °C (272.2 °F) and a working chamber pressure
that meets Supervision Regulation on Safety Technology for Fixed Type Pressure Vessel.
With years of research and customer feedback we have fine-tuned our formula for perfectly
optimized cycles with the precise combination of pulses, timing, temperature control, speed
and drying.
Applications
Bedspreads, linens, laundry, medical waste, etc.. General-purpose steam sterilization of
surgical instruments, textiles and hospital utensils in central sterilization departments,
operation departments and laundries.
Competitive Advantage
Medpharm strong engineering capability allows for highly customized autoclave design.
Facilities with space layout limitations will benefit from custom space design autoclaves which
provide high capacity in a limited space.
ZHANGJIAGANG MEDPHARM MACHINERY LTD.
MEDPHARM Company team is committed to providing quality sterilization products, service, parts and
maintenance training at significant savings. From customized parts packages to turnkey equipment
acquisition and installation, our innovative sterilization solutions are designed to meet our customers' needs.
We understand that untimely breakdowns are costly and can be detrimental to your everyday operations.
Our service team is compromised of expert technicians that respond quickly to all of your service needs.
We offer on-site service support including inservices, installation and deinstallation of your equipment. From
breakdown repairs to comprehensive maintenance programs, we have an innovative service solution for you.
We will service on-Site Repairs.Our staff is equipped with the resources required to provide you with
on-site, on-time repairs in compliance with manufacturer specifications. We also provide installation,
deinstallation, and inservices on most makes and models of sterilization equipment. Near or far we can
provide you with service support that you need.
Our reconditioning background has taught us of the importance of having quality parts on hand at all times.
We have a large inventory of individual repair parts, as well as preventive maintenance kits, and consumables
in stock and ready to ship today.
With our expert technical support staff, it is not always necessary to have part numbers to place an order.
We will help you identify the quality parts that you need.